Exposure - Wrist

From wikiRadiography
Jump to navigation Jump to search
This page contains radiographic exposure values

Adult Wrist

Region View kVp mAs Cassette Grid Distance
WristPA522.5DetailYes100 - 115cm (40 inches)

PA Oblique (Lateral Rotation) 523.2DetailYes100 - 115cm (40 inches)

PA Oblique (Medial Rotation)523.2DetailYes100 - 115cm (40 inches)

Lateral534DetailYes100 - 115cm (40 inches)

Ulna Deviation522.5DetailYes100 - 115cm (40 inches)

Radial Deviation522.5DetailYes100 - 115cm (40 inches)

Carpal Tunnel502DetailYes100 - 115cm (40 inches)

Carpal Bridge502DetailYes100 - 115cm (40 inches)

Paediatric Wrist

Region View kVp mAs Cassette Grid Distance
WristPA502DetailYes100 - 115cm (40 inches)

PA Oblique (Lateral Rotation) 502.5DetailYes100 - 115cm (40 inches)

Lateral523.2DetailYes100 - 115cm (40 inches)


These are average exposures using a Carestream CR system, exposures may vary dramatically between different film/screen combinations, CR or DR systems

Exposure Chart....Go Back to Exposure Values homepage