Region View kVp mAs Cassette Grid Distance
Thoracic AP6620RegularYes100 - 115cm (40 inches)

Lateral (Breathing technique)6650 x 2.5 secondsRegularYes100 - 115cm (40 inches)

Oblique6650RegularYes100 - 115cm (40 inches)

Pediatric Thoracic Spine

Region View kVp mAs Cassette Grid Distance
ThoracicAP6016RegularYes100 - 115cm (40 inches)

Lateral (Breathing technique)6030 x 2.5 secondsRegularYes100 - 115cm (40 inches)

Oblique6630RegularYes100 - 115cm (40 inches)


These are average exposures using a Carestream CR system, exposures may vary dramatically between different film/screen combinations, CR or DR systems