Exposure - Coccyx

From wikiRadiography
Jump to navigation Jump to search
This page contains radiographic exposure values

Adult Coccyx

Region View kVp mAs Cassette Grid Distance
CoccyxAP6520RegularYes100 - 115cm (40 inches)

Lateral 7030RegularYes100 - 115cm (40 inches)

Paediatric Coccyx

Region View kVp mAs Cassette Grid Distance
CoccyxAP6016RegularYes100 - 115cm (40 inches)

Lateral 6520RegularYes100 - 115cm (40 inches)


These are average exposures using a Carestream CR system, exposures may vary dramatically between different film/screen combinations, CR or DR systems

Exposure Chart....Go Back to Exposure Values homepage


Coccyx Links

Coccyx Positioning
Coccyx Radiographic Anatomy
  • Coccyx - Adult
  • Coccyx - Paediatric
Coccyx Miscellaneous